Dimensions and Structure

Barometric Absolute Pressure Transmitter
Absolute atmospheric pressure transmitter for environmental monitoring, small weather stations, altitude-related instrumentation and building applications. The compact unit measures non-condensing air over 15–130 kPa and provides a selectable current, voltage or Modbus RTU output, with an optional local LCD.
Output and Electrical Selection
| Output code | Signal | Wiring | Supply | Output load |
|---|---|---|---|---|
| A4 | 4–20 mA | Two-wire | 12–30 VDC | ≤250 Ω |
| V10 | 0–10 VDC | Three-wire | 12–30 VDC or isolated 24 VAC ±20% | >5 kΩ |
| RS | RS485, Modbus RTU | Four-wire | 12–30 VDC or isolated 24 VAC ±20% | Not specified |
Display Selection
| Code | Configuration |
|---|---|
| N | No display |
| O | LCD display; display area45×30 mm |
Operating Specifications
| Item | Value |
|---|---|
| Pressure reference | Absolute atmospheric pressure |
| Medium | Non-condensing air |
| Measurement range | 15–130 kPa |
| Accuracy | ±0.25% FS |
| Resolution | 0.01 kPa |
| Allowable overpressure | 200 kPa |
| Operating temperature | −20 to +70 °C |
| Compensation temperature | −10 to +60 °C |
| Storage temperature | −40 to +70 °C |
| Protection | IP65 |
| Communication | RS485 standard interface, Modbus RTU, with RS output option |
Dimensions
| Width (mm) | Overall height (mm) | Body height (mm) | Depth (mm) | Mounting span (mm) | Probe diameter (mm) | Cable gland |
|---|---|---|---|---|---|---|
| 100 | 110 | 80 | 46 | 88 | 15 | M16 |
Catalog reference:LFM57. Ordering structure:series – display – output; for example LFM57-O-A4 identifies the display version with4–20mA output. Select supply and load for the chosen output. Use an isolated AC supply for AC-powered voltage or RS485 versions. The200kPa overpressure rating is not the measurement range. Dimensions are in millimetres; the original drawing label identifies the M16 waterproof cable gland.