DODOMACH PNEUMATIC

Pressure Transmitters

Combined Pressure and Temperature Transmitter

Compact combined transmitter with a piezoresistive pressure sensor and Pt100 temperature sensor in one housing. It monitors both pressure and temperature in compatible gas or liquid systems, with selected current, voltage or RS485 output configurations for water treatment and industrial process monitoring.

Dimensions and Structure

Combined Pressure and Temperature Transmitter dimensions

Combined Pressure and Temperature Transmitter

Compact combined transmitter with a piezoresistive pressure sensor and Pt100 temperature sensor in one housing. It monitors both pressure and temperature in compatible gas or liquid systems, with selected current, voltage or RS485 output configurations for water treatment and industrial process monitoring.

Measurement Specifications

Item Value
Measured variables Pressure and temperature
Pressure sensor Piezoresistive
Temperature sensor Pt100
Pressure range selection Zero-based spans with full scale from 10 kPa to 70 MPa
Temperature measurement range −20 to +100 °C
Pressure accuracy ±0.5% FS
Temperature accuracy ±2% FS
Accuracy basis At 25 °C; includes linearity, repeatability and hysteresis
Overload 1.5 times rated pressure
Stability <0.5% FS/year
Working temperature −20 to +85 °C
Pressure reference Gauge (G) or absolute (A)
Medium Gases or liquids compatible with 304/316L stainless steel and selected FKM or NBR seals

Output and Supply Options

Code Signal option Supply
A4 4–20 mA 10–30 VDC
V05 0.5–4.5 V 4.75–5.25 VDC
RS RS485 10–30 VDC

Electrical Connections

Code Connection Catalog protection
D1 DIN43650A angular connector IP65
M M12 waterproof cable outlet IP65
C3 GX12 three-core connector IP54
C4 GX12 four-core connector IP54
H M12 four-core connector IP65

Process Connections

Thread Catalog selection code
G1/4 G
G1/2 G2
NPT1/4 N
M20×1.5 M20
M10×1 Confirm code for selected catalog revision
M12×1 M12

Other Selection Fields

Field Selection
Pressure unit K: kPa; M: MPa; P: psi; B: bar
Pressure accuracy code 0.5
Temperature accuracy code 2
Cable length 1.0: 1 m; 2.0: 2 m; 3.0: 3 m where applicable

Catalog reference: LFT3100. Select both measuring ranges, output configuration, accuracy, electrical connector, pressure connection and cable length. Confirm the pressure and temperature channel assignments and complete wiring diagram for the selected connector before connection; the output table is not a universal pinout. Intelligent RS485 versions support Modbus according to the manufacturer; confirm the protocol settings and register map for the delivered unit. The original drawings show multiple enclosure variants, with dimensions in millimetres; confirm the relevant drawing for the ordered configuration. The photograph shows one DIN43650A version. Confirm medium and storage temperature limits separately because source editions use different labels. No hazardous-area certification or international-brand interchangeability is claimed.