DODOMACH PNEUMATIC

Pressure Transmitters

Barometric Absolute Pressure Transmitter

Absolute atmospheric pressure transmitter for environmental monitoring, small weather stations, altitude-related instrumentation and building applications. The compact unit measures non-condensing air over 15–130 kPa and provides a selectable current, voltage or Modbus RTU output, with an optional local LCD.

Dimensions and Structure

Barometric pressure transmitter dimensions

Barometric Absolute Pressure Transmitter

Absolute atmospheric pressure transmitter for environmental monitoring, small weather stations, altitude-related instrumentation and building applications. The compact unit measures non-condensing air over 15–130 kPa and provides a selectable current, voltage or Modbus RTU output, with an optional local LCD.

Output and Electrical Selection

Output code Signal Wiring Supply Output load
A4 4–20 mA Two-wire 12–30 VDC ≤250 Ω
V10 0–10 VDC Three-wire 12–30 VDC or isolated 24 VAC ±20% >5 kΩ
RS RS485, Modbus RTU Four-wire 12–30 VDC or isolated 24 VAC ±20% Not specified

Display Selection

Code Configuration
N No display
O LCD display; display area45×30 mm

Operating Specifications

Item Value
Pressure reference Absolute atmospheric pressure
Medium Non-condensing air
Measurement range 15–130 kPa
Accuracy ±0.25% FS
Resolution 0.01 kPa
Allowable overpressure 200 kPa
Operating temperature −20 to +70 °C
Compensation temperature −10 to +60 °C
Storage temperature −40 to +70 °C
Protection IP65
Communication RS485 standard interface, Modbus RTU, with RS output option

Dimensions

Width (mm) Overall height (mm) Body height (mm) Depth (mm) Mounting span (mm) Probe diameter (mm) Cable gland
100 110 80 46 88 15 M16

Catalog reference:LFM57. Ordering structure:series – display – output; for example LFM57-O-A4 identifies the display version with4–20mA output. Select supply and load for the chosen output. Use an isolated AC supply for AC-powered voltage or RS485 versions. The200kPa overpressure rating is not the measurement range. Dimensions are in millimetres; the original drawing label identifies the M16 waterproof cable gland.